Calibration of critical flow nuzzles by volumetric method Cd

the present work, the objective is to improve the measurement of gas flow through the use of critical flow nozzles (Critical Flow Venturi Nozzles), for this, the calibration procedure of two critical flow nozzles with diameters in the throat of 0.022 and 0.088 inches. The procedure developed is base...

ver descrição completa

Na minha lista:
Detalhes bibliográficos
Principais autores: Rivera López, Jesús Eduardo, Díaz Martínez, Alejandro, Tamayo Mesa, Pedro Alejandro, Gutiérrez Paredes, Guadalupe Juliana, Orozco Durán, Gabriela Esmeralda, Arciniega Martínez, José Luis
Formato: Online
Idioma:spa
Publicado em: Universidad Autónoma de Baja California 2021
Acesso em linha:https://recit.uabc.mx/index.php/revista/article/view/139
Tags: Adicionar Tag
Sem tags, seja o primeiro a adicionar uma tag!
Descrição
Resumo:the present work, the objective is to improve the measurement of gas flow through the use of critical flow nozzles (Critical Flow Venturi Nozzles), for this, the calibration procedure of two critical flow nozzles with diameters in the throat of 0.022 and 0.088 inches. The procedure developed is based on the recommendations made by the ISO 9300 Standard for the estimation and calculation of the discharge coefficient "Cd" of the critical flow nozzles. The calibration was carried out at the National Metrology Center (CENAM) with the use of national gas flow standards. The standards used or testers are of the type: Piston and Bell, with a measurement range of 0.05 to 50 L / min, and from 50 to 2840 L / min, respectively. The method used to calibrate the nozzles was by direct comparison with the standard.